Model: HY-AsT1000

HYTEZ HY-AsT1000 Online Arsenic Total Analyzer fully automated the Arsenic Molybdenum Blue Spectrophotometry Method.

Features

  • Reliable, Exact and Fully Automatic Analysis
  • User Friendly Software Interface
  • Touch Screen
  • Self Calibrating and Monitoring
  • High Measuring Accuracy
  • Easy Maintenance
  • Minimal Reagent and Sample Consumption
  • Automatic Discharge Residual in analyzer and resume working status when Power resume after abnormal power reset

Applications

  • Industrial Waste Water
  • Surface Water

Technical Specification

Measurement Method Arsenic Molybdenum Blue Spectrophotometry Method
Measuring Range0 – 3.00 mg/L as As(Total)
0.05 – 5.00 mg/L as As(Total)
0.1 – 10.00 mg/L as As(Total)
Accuracy< 0.2 mg/L for As(Total) Reading < ± 0.02 mg/L
≥ 0.2 mg/L as As(Total) Reading < ± 10%
Repeatability± 5%
Measurement PeriodMinimum Measurement Period: 35 Minutes (According to Actual Water Sample)
[Digestion Time can be modified from 10 ~ 120 Minutes)
Sampling ModeTime Interval Measurement Mode: 55 ~ 9999 Minutes (Adjustable)
Integral Time Measurement Mode: eg: at 0100hr, 0300hr etc
Remote Start Mode: Require Control Display Module
Calibration ModeAutomatic Mode: 1 ~ 99 Days (Adjustable)
Manual Mode
Outputs1 x 4 ~ 20mA (Arsenic Total Reading)
RS232 or RS485 Communication Mode
Relays Output
Ambient ConditionsIndoor or Enclosure Installation
Temperature: 5 ~ 40 deg C
Power230VAC ±10% V,50±10% Hz,5 A
Dimension450mm x 510mm x 1450mm (L x W x H)
Subject to Change without Notice